Device measurement averages taken across all six patient plans. The measurement used for the reproducibility analysis is given in parentheses (percent of pixels passing gamma for the planar devices, the nC charge reading for the cc04 ion chamber, and the dose (cGy) measured for the multiple ion chamber (MIC) phantom. The averages reported are for both the “redelivery” and “total delivery” measurements (three measurements each), which over the six patients translates to averages taken over 18 measurements each for “redelivery” and “total delivery”
Arccheck Helical Diode Array Sun Nuclear, supplied by Sun Nuclear Corporation, used in various techniques. Bioz Stars score: 90/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/product/arccheck+helical+diode+array+sun+nuclear/pmc04048867-48-33-37?v=Sun+Nuclear+Corporation
Average 90 stars, based on 1 article reviews
arccheck helical diode array sun nuclear - by Bioz Stars, 2026-07
90/100 stars

Images

1) Product Images from "Reproducibility in patient‐specific IMRT QA"

Article Title: Reproducibility in patient‐specific IMRT QA

Journal: Journal of Applied Clinical Medical Physics

doi: 10.1120/jacmp.v15i3.4741

Device measurement averages taken across all six patient plans. The measurement used for the reproducibility analysis is given in parentheses (percent of pixels passing gamma for the planar devices, the nC charge reading for the cc04 ion chamber, and the dose (cGy) measured for the multiple ion chamber (MIC) phantom. The averages reported are for both the “redelivery” and “total delivery” measurements (three measurements each), which over the six patients translates to averages taken over 18 measurements each for “redelivery” and “total delivery”
Figure Legend Snippet: Device measurement averages taken across all six patient plans. The measurement used for the reproducibility analysis is given in parentheses (percent of pixels passing gamma for the planar devices, the nC charge reading for the cc04 ion chamber, and the dose (cGy) measured for the multiple ion chamber (MIC) phantom. The averages reported are for both the “redelivery” and “total delivery” measurements (three measurements each), which over the six patients translates to averages taken over 18 measurements each for “redelivery” and “total delivery”

Techniques Used:

Venn diagram showing statistically significant grouping of dosimetric systems based on their “total delivery” reproducibility. The coefficient of variation is given in parentheses. The coefficient of variation of film was significantly different from the cc04 ion chamber, MapCHECK with AP beams formed into a composite plane, MapCHECK with AP beams analyzed in a field‐by‐field analysis, and the multiple ion chamber (MIC) phantom. The coefficient of variation of the ArcCHECK and the MapCHECK with the original rotational gantry angles were not significantly different from either group.
Figure Legend Snippet: Venn diagram showing statistically significant grouping of dosimetric systems based on their “total delivery” reproducibility. The coefficient of variation is given in parentheses. The coefficient of variation of film was significantly different from the cc04 ion chamber, MapCHECK with AP beams formed into a composite plane, MapCHECK with AP beams analyzed in a field‐by‐field analysis, and the multiple ion chamber (MIC) phantom. The coefficient of variation of the ArcCHECK and the MapCHECK with the original rotational gantry angles were not significantly different from either group.

Techniques Used:

Isolating the error caused by setup alone from the redelivery and total coefficients of variation (CV) averaged across all patient plans
Figure Legend Snippet: Isolating the error caused by setup alone from the redelivery and total coefficients of variation (CV) averaged across all patient plans

Techniques Used:



Similar Products

90
Sun Nuclear Corporation arccheck helical diode array sun nuclear
Device measurement averages taken across all six patient plans. The measurement used for the reproducibility analysis is given in parentheses (percent of pixels passing gamma for the planar devices, the nC charge reading for the cc04 ion chamber, and the dose (cGy) measured for the multiple ion chamber (MIC) phantom. The averages reported are for both the “redelivery” and “total delivery” measurements (three measurements each), which over the six patients translates to averages taken over 18 measurements each for “redelivery” and “total delivery”
Arccheck Helical Diode Array Sun Nuclear, supplied by Sun Nuclear Corporation, used in various techniques. Bioz Stars score: 90/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/product/arccheck+helical+diode+array+sun+nuclear/pmc04048867-48-33-37?v=Sun+Nuclear+Corporation
Average 90 stars, based on 1 article reviews
arccheck helical diode array sun nuclear - by Bioz Stars, 2026-07
90/100 stars
  Buy from Supplier

Image Search Results


Device measurement averages taken across all six patient plans. The measurement used for the reproducibility analysis is given in parentheses (percent of pixels passing gamma for the planar devices, the nC charge reading for the cc04 ion chamber, and the dose (cGy) measured for the multiple ion chamber (MIC) phantom. The averages reported are for both the “redelivery” and “total delivery” measurements (three measurements each), which over the six patients translates to averages taken over 18 measurements each for “redelivery” and “total delivery”

Journal: Journal of Applied Clinical Medical Physics

Article Title: Reproducibility in patient‐specific IMRT QA

doi: 10.1120/jacmp.v15i3.4741

Figure Lengend Snippet: Device measurement averages taken across all six patient plans. The measurement used for the reproducibility analysis is given in parentheses (percent of pixels passing gamma for the planar devices, the nC charge reading for the cc04 ion chamber, and the dose (cGy) measured for the multiple ion chamber (MIC) phantom. The averages reported are for both the “redelivery” and “total delivery” measurements (three measurements each), which over the six patients translates to averages taken over 18 measurements each for “redelivery” and “total delivery”

Article Snippet: Four commercial dosimeters and one in‐house‐designed dosimeter were used for this study: EDR2 radiographic film (Kodak Carestream, Rochester, NY), Wellhofer cc04 ion chamber (CNMC, Nashville, TN), MapCHECK2 2D diode array (Sun Nuclear Corporation), ArcCHECK helical diode array (Sun Nuclear Corporation), and an in‐house‐designed multiple ion chamber (MIC) phantom.

Techniques:

Venn diagram showing statistically significant grouping of dosimetric systems based on their “total delivery” reproducibility. The coefficient of variation is given in parentheses. The coefficient of variation of film was significantly different from the cc04 ion chamber, MapCHECK with AP beams formed into a composite plane, MapCHECK with AP beams analyzed in a field‐by‐field analysis, and the multiple ion chamber (MIC) phantom. The coefficient of variation of the ArcCHECK and the MapCHECK with the original rotational gantry angles were not significantly different from either group.

Journal: Journal of Applied Clinical Medical Physics

Article Title: Reproducibility in patient‐specific IMRT QA

doi: 10.1120/jacmp.v15i3.4741

Figure Lengend Snippet: Venn diagram showing statistically significant grouping of dosimetric systems based on their “total delivery” reproducibility. The coefficient of variation is given in parentheses. The coefficient of variation of film was significantly different from the cc04 ion chamber, MapCHECK with AP beams formed into a composite plane, MapCHECK with AP beams analyzed in a field‐by‐field analysis, and the multiple ion chamber (MIC) phantom. The coefficient of variation of the ArcCHECK and the MapCHECK with the original rotational gantry angles were not significantly different from either group.

Article Snippet: Four commercial dosimeters and one in‐house‐designed dosimeter were used for this study: EDR2 radiographic film (Kodak Carestream, Rochester, NY), Wellhofer cc04 ion chamber (CNMC, Nashville, TN), MapCHECK2 2D diode array (Sun Nuclear Corporation), ArcCHECK helical diode array (Sun Nuclear Corporation), and an in‐house‐designed multiple ion chamber (MIC) phantom.

Techniques:

Isolating the error caused by setup alone from the redelivery and total coefficients of variation (CV) averaged across all patient plans

Journal: Journal of Applied Clinical Medical Physics

Article Title: Reproducibility in patient‐specific IMRT QA

doi: 10.1120/jacmp.v15i3.4741

Figure Lengend Snippet: Isolating the error caused by setup alone from the redelivery and total coefficients of variation (CV) averaged across all patient plans

Article Snippet: Four commercial dosimeters and one in‐house‐designed dosimeter were used for this study: EDR2 radiographic film (Kodak Carestream, Rochester, NY), Wellhofer cc04 ion chamber (CNMC, Nashville, TN), MapCHECK2 2D diode array (Sun Nuclear Corporation), ArcCHECK helical diode array (Sun Nuclear Corporation), and an in‐house‐designed multiple ion chamber (MIC) phantom.

Techniques: